MRI Physics, Part I: Interactive Companion Experimental

Teaching page. A companion to Ridgway 2010, read alongside the paper. The drawings are schematic models, not patient images; values marked illustrative are model values, not reference values.

System and signal

The three magnetic fields of the scanner

The main magnet makes a strong, constant field, B0, along the bore (the z axis). One tesla is about 20,000 times the earth's field; clinical systems run from 0.2 to 3.0 T, and 1.5 T is the most common for cardiac work.

Three gradient coils add a weaker field in the same direction as B0 whose strength changes with position along x, y or z; their steepness is given in mT/m, and they switch on and off rapidly. The rf transmitter coil, closest to the patient, makes the much smaller B1 field, oscillating in the megahertz range. A separate receiver coil over the heart picks up the signal.

Read in the paper: Ridgway 2010, section “MR system components”, Figure 1.

The field along x is B0 + G × x; each position then has its own Larmor frequency (next section). Gradient strengths here are illustrative; the paper gives the unit, not a value.

Origin of the MR signal: spins, net magnetisation and the Larmor frequency

The signal comes from hydrogen nuclei (single protons) in water and fat. Each proton's spin makes a tiny magnetic moment. In B0 the moments line up toward or against the field, with a small excess along it, typically just a few per million; the excess adds up to the net magnetisation, M0, along +z. A stronger B0 gives a bigger excess and a bigger M0.

Energy is delivered at the Larmor frequency, ω0 = γ × B0, where the gyromagnetic ratio γ is 42.6 MHz/T for the proton: about 64 MHz at 1.5 T.

Read in the paper: Ridgway 2010, section “Origin of the MR signal”, Figure 2a.

The excess drawn here is exaggerated tens of thousands of times to be visible (up to 12 spins in 60); the real excess is a few per million.

Radiofrequency pulses and flip angle

Switched on at the Larmor frequency, the rf field B1 rotates with the precessing magnetisation, so M tips away from z while it precesses: a spiral path. The angle reached when the pulse stops is the flip angle; it grows with the pulse's amplitude and duration.

M then has a longitudinal part, Mz, and a transverse part, Mxy; only the rotating Mxy gives a signal. A 90° pulse (saturation) puts all of M into the xy plane, the largest signal from one pulse. A low flip angle (α, e.g. 30°) leaves some Mz, so it can be repeated quickly. A 180° inversion pulse at equilibrium turns M to −z: no signal, a preparation for inversion recovery and black blood.

Read in the paper: Ridgway 2010, section “Radiofrequency pulses and flip angle”, Figure 2b–e.

Drawn in the laboratory frame with precession slowed to a few turns per pulse; real precession runs at the Larmor frequency, about 64 MHz at 1.5 T.

A section-by-section companion to Ridgway 2010, part I: each section condensed, with a model you can move. The drawings are schematic simulations, not patient images.

Method and sources

What this page is

A teaching companion to Ridgway 2010, in the paper's order. Each section is condensed and paraphrased in our own words, and each of the paper's figures that explains a mechanism is redrawn as an interactive model. Nothing here is a clinical threshold. The constants and ranges the page prints (the gyromagnetic ratio, field strengths, Table 1 timings, echo train lengths, the black-blood TI and slab values) are the paper's; every other parameter is an illustrative model value, labelled where it is used.

Models

Left out

The paper's reference list, its vendor names for non-cine sequences beyond those shown, and parallel imaging, which it mentions only in passing (part II of the series covers it). Image acquisition time is TR × NP as printed; averages are not in this paper's formula and are not added.

Verification

Experimental. The pure functions (Larmor frequency, relaxation, echo timing, slice thickness, the Fourier round trip, acquisition time, washout, low flip angle, inversion null, cine arithmetic) are unit-tested against hand-checked values; every sentence the page writes about a simulation is computed from the model at the current setting. The models are not verified against scanner data.

Source and licence

  1. Ridgway JP. Cardiovascular magnetic resonance physics for clinicians: part I. J Cardiovasc Magn Reson 2010;12:71. doi:10.1186/1532-429X-12-71

The paper is open access under the Creative Commons Attribution License 2.0 (CC BY 2.0), © 2010 Ridgway. This page adapts it: the text is condensed and paraphrased, and the figures are redrawn as interactive models; no figure is copied. The adaptation is ours, not the author's.