HeartCalc / Neonatal echo ref

Neonatal Echo Reference ValuesExperimental

Published preterm and term echo reference ranges and cut-offs, each with its population, method and source

Key values only. Each row gives the main ranges or cut-offs from one source and links to the full paper. Populations, ages and methods differ between rows, so values are not interchangeable; “cut-off” rows grade disease and are not normal ranges.

Reference table

Neonatal echo reference values and cut-offs
ParameterValuesPopulationHow measuredSource

RVSP from the tricuspid regurgitation jet

Simplified Bernoulli, as given by de Boode et al. 2018 (Eq. 3): RVSP = 4 × (TR Vmax)² + RAP. Enter the peak TR velocity in m/s and the right atrial pressure you assume or measured; systolic blood pressure is optional and gives RVSP as a fraction of systemic pressure.

TR Vmax: continuous-wave Doppler aligned with the jet, best envelope. RAP has no default here; state the value you assume. A poorly aligned or incomplete jet underestimates RVSP; no TR jet does not mean normal pressure.

Model, equations and sources

How the values were chosen

RVSP equation

p = 4 × v²                      (p mm Hg, v m/s)
SPAP ≈ RVSP = 4 × (TR Vmax)² + RAP

de Boode WP, Singh Y, Molnar Z, et al. Application of Neonatologist Performed Echocardiography in the assessment and management of persistent pulmonary hypertension of the newborn. Pediatr Res 2018;84(Suppl 1):68-77. doi:10.1038/s41390-018-0082-0, Eq. 2 and 3 (open-access full text). The ratio to systolic blood pressure is a simple division shown for context; no category is assigned to it.

Not included or not verified

Computed in this browser; nothing you enter is stored or sent. Experimental: last checked against the cited sources October 2026.

Reference values describe study populations. They do not replace clinical judgement or the full published tables.