| Too dark or too bright |
- Overall gain: just a few echoes in the blood, crisp blood–endocardial borders
- TGC: even brightness from top to bottom; uneven pods cause bands
- Dynamic range (compression), see the next problem
- Brighten with receiver gain before output power; raise power only when the signal is weak against noise (educational site)
|
In 3D, dropout from low gain or uneven TGC cannot be recovered later |
ASE 2019; EAE/ASE 2012; echocardiographer.org |
| Haze or noise in a cavity |
- Lower the gain
- Harmonic imaging (fewer side-lobe echoes)
- Focus at the depth of interest
- Change the window or angle
|
Near-field clutter can mimic an apical LV thrombus: it ignores wall motion and crosses the wall; check with colour at a low scale, other planes, contrast, CT or CMR |
ASE 2026 §3.2, 3.3, 9 |
| Blurry, little detail |
- Highest transmit frequency that still reaches the structure
- Focus at the structure
- Less depth; structure centred
- Acquisition (preprocessing) zoom, not zoom on the stored image
- Window closest to the structure; measure along the beam
|
Higher frequency penetrates less. Lateral resolution falls with depth; axial is better than lateral |
ASE 2019; ASE 2024 pediatric |
| Flat gray or harsh black and white |
- Dynamic range (compression): lower for a poor window (more contrast), higher to see subtle tissue (more grays)
- The lab’s gray map, kept the same for serial studies
|
Too few grays can hide a thin wall, thrombus or vegetation; too many look washed out. In coronaries, extra compression may hide a luminal thrombus |
ASE 2019; ASE 2024 pediatric |
| Jerky, low frame rate |
- Less depth
- Narrower sector
- One focal zone
- Acquisition zoom
- Fewer scan lines, where the system allows
- Persistence (frame averaging) low for the heart
|
Fewer lines or a narrower sector cost lateral resolution or field of view. Children’s faster heart rates need higher frame rates than adults’ |
ASE 2019; ASE 2024 pediatric; Nestaas 2018; Zander 2020 (background) |
| Ghost or double structure |
- Change the window or angle
- Lower the gain
- Harmonic imaging (side lobe)
- Focus at the depth of interest; higher frequency (beam width)
|
Reverberation ignores anatomic boundaries; an artifact moves with the structure that makes it, not on its own. If unresolved, use 3D, contrast, CT or CMR |
ASE 2026 §2–3 |
| Shadow or missing wall |
- Image from a window where the structure is perpendicular to the beam
- Change the plane or angle around a shadowing reflector
- Colour Doppler to tell a true defect from dropout
|
A septum parallel to the beam can drop out falsely; use more than one view |
ASE 2024 pediatric; ASE 2026 §2.5, 6.2, 9, Table 2 |
| Colour bleeds past the flow |
- Colour gain: raise until speckle appears outside the flow, then lower until it just disappears
- Raise the wall filter (blooming)
|
The same jet looks larger at a lower Nyquist and higher colour gain; record blood pressure with jet size |
ASE 2019; ASE 2026 §5.7 |
| Unwanted colour aliasing |
- Raise the scale (Nyquist)
- Shift the baseline
- Less depth
- Lower transmit frequency
|
Aliasing is used on purpose for PISA (baseline shifted toward the jet) |
ASE 2026 §5.6, 10 |
| No colour in slow flow |
- Lower the Nyquist to about 30 cm/s for atria and pulmonary veins Adult
- Low wall filter
- Colour gain up to just below noise
|
Routine jets use 50–70 cm/s Adult. No pediatric guideline gives a number |
ASE 2019; ASE 2002 |
| Colour frame rate low |
- 2D sector shallow and narrow first
- Smallest colour box that covers the flow
- Side-by-side colour compare only when needed
|
In children keep the colour frame rate ≥20 Hz |
ASE 2024 pediatric Table 7; ASE 2019 |
| Jet in an odd place |
- Another window or angle
- Lower colour gain
- Adjust the focus and sweep for an out-of-plane jet
|
Suspect an artifact when a jet has no flow convergence or vena contracta |
ASE 2026 §5.1–5.5 |
| Spectral aliasing |
- Raise the scale
- Shift the baseline away from the flow
- Shallower sample (higher PRF)
- Lower transmit frequency
- High PRF or CW
|
CW and high PRF cannot say where along the beam the velocity is; some systems switch to high PRF when the scale is raised |
ASE 2026 §4.3; ASE 2019 |
| Weak or incomplete envelope |
- Align with flow, colour first
- A little more spectral gain
- Nonimaging CW probe
- Other windows
|
Measure only a well-defined envelope; not faint signal outside it |
ASE 2002; ASE 2019 |
| Velocity looks too high |
- Lower the spectral gain
- Measure the modal (densest) envelope, outer edge
- At a CW peak, ignore the fine linear “beard”
- If two jets overlap, use a smaller sample, avoid far field, other window
|
Overgain overestimates velocity; so does measuring contrast blooming |
ASE 2026 §4.1, 4.5, 4.6, 7.3; ASE 2002 |
| Velocity looks too low |
- Align with flow from several windows (nonimaging probe)
- Report the highest-velocity window; don’t average windows
- No angle correction
|
Underestimation 6% at 20°, 50% at 60°; the gradient error is larger (4V²) |
ASE 2026 §4.4; ASE 2019; ASE 2002 |
| Clutter or gap at the baseline |
- Wall filter: show the start and end of the flow
- Low for slow flow (for example pulmonary veins); higher for fast jets
|
A high filter removes the slow parts of the signal |
ASE 2019; ASE 2026 §4.2 |
| Tissue Doppler looks wrong |
- Tissue Doppler preset
- Low gain
- Small sample at the annulus; neonates 2 mm
- Beam parallel to the wall
- Neonates: scale ±0.16 m/s; freeze the 2D while recording
|
High gain overestimates pulsed TDI. Colour TDI reads about 20% lower than pulsed; not interchangeable |
ASE 2024 neonatal; Nestaas 2018 |
| Strain or 3D to follow |
- Frame rate: targets on Strain; narrow sector, less depth
- Same vendor and software for follow-up
- 3D: optimise 2D first; gain and compression mid-range, slightly higher TGC Adult; breath-hold; fewest beats
|
Strain from copies stored at 30 frames/s differed from acquisition-rate images abstract |
ASE 2024 pediatric; ASE/EACVI 2025; EAE/ASE 2012; Koopman 2011 |